HOME > Power System > Measuring Instrument > Energy Meter Test Bench > Energy Meter Test Bench PTC 8320D

PTC 8320D three phase energy auto-calibration system is a new generation, high precision, energy meter calibration equipment. It uses the advanced technology and components available in update technology, follows the national standard and regulation to develop and design. This bench can be used to test both single / 3 phase energy meter (up to class 0.2%). This bench is using the aluminum compound configuration to form with digital controlled signal source with PWM amplifier, HC3100 standard precision instrument, error processor system and auto-testing software.
Energy Meter Test Bench PTC 8320D

 

Features
  • PWM amplifier provide high ranges output of voltage and current, the power source efficiency better than 85%.
  • Auto-range for current / voltage.
  • Overload protection for voltage / current.
  • Auto / manual test by keyboard/software on bench.
  • Come with the 3 phase precision energy meter (0.05% class) with RS232 communication port (Model : HC 3100).
  • Suitable for different single* / 3 phase electronic / interaction energy meter.
  • Suitable for 4 phase test.
  • Calibration software with PC based under Windows 98 / 2000 / XP which the result can be saved, searched, and printed out.
  • Database provides with statistics, category, information check, management and online function.

* OPTION: For test many single meter in PTC 8320D in one time, must add a single MSVT.

Model & Ordering

      MODEL : PTC 8320D

Specification
  • Accuracy class :
    • 0.5% class
    • 0.05% on requested
    • 0.1% on requested
  • Standard power testing equipment : HC 3100 three phase multi-functions precision power testing instrument......
    • Accuracy class : 0.05% class
    • Input range :
      • 3 * 60V / 120V / 240V / 420V auto range
      • 3 * 1A / 10A / 100A auto range
  • Output range :
    • Output voltage : 3 * 60V / 120V / 240V / 420V
    • Output current : 3 * 0.1 ~ 120A
  • Maximum output capacity of each phase :
    • Voltage : 100VA (To be increased according to the number of meters calibrated)
    • Current : 200VA (120A) (To be increased according to the number of meters calibrated)
  • Adjustable range :
    • Voltage : 0 ~ 120%, Fine adjustment : 0.01%
    • Current : 0 ~ 120A, Fine adjustment : 0.0001A (minimum)
    • Phase adjistable (Voltage / current) : 0 - 360.0 , Fine adjustment : 0.1
    • Frequency : 45 ~ 65Hz, Fine adjustment : 0.01Hz
  • Output Stability of Power factor < 0.1% (100s)
  • Waveform distortion coefficient : Voltage / current < 0.5%
  • Harmonic wave output :
    • 2 ~ 21 sub-harmonic wave Contents of Harmonic Wave < 40%
    • 3, 5, 7 sub-harmonic wave coupling, contents of each harmonic wave <10%
    • Phase angle control (odd sub-harmonic wave)
    • Wave group control (sub-harmonic wave)
  • Starting current output : 2mA (min)
  • Starting current / power factor measurement : 5% (2mA)
  • Numbers of cailbrated energy meters : 6 meters (8,12,16 on requested)
  • Comm. port : RS485 or on requested (option)
  • Test function :
  •     Basic error test
        Typical deviation (S) in error test
        Initial test / automatic / manual operation
        Starting test / automatic / manual operation
        Phase rotation test
        Deviation test for 24 hours
       *Demand error test
       *Clock error test
       *Harmonic test
       *Burden test
       *RS485 communication port test
      Note : "*" means "Optional function".
  • Input power : 3 * 220V / 380V + 10%, 50Hz, max. current : 3 * 10A
  • Dimension : 2370 (W) * 1505 (H) * 750 (D) / mm
  • OPTION: MSVT (Multi Secondary Voltage Transformer)
  • Specification
    • Primary voltage : 180 ~ 300 V (Normal Voltage : 240V)
    • Secondary voltage : 180 ~ 300V
    • Secondary VA rating : 15VA per winding
    • Insulation secondary to secondary : 500V, 50Hz, 60 seconds
    • Primary to secondary ratio accuracy : + 0.1% (typical)
    • Secondary to secondary ratio accuracy less than : + 0.02%
    • Secondary to secondary phase angle accuracy less than + 1 minute
    • Number of secondary output : 6 channel (option up to 60 channel)