Intelligent Power Instrument
The main configuration of PTC 9300H is a computerized 3 phase power source with built-in precision reference instrument. It sets new industry standards in the field of test & calibration measurements. Basically, PTC 9300H fulfills 3 important functions :
Reference output readings
Power transducers' calibrations
Power meter calibrations
Test data and setting ranges can be printed out via the parallel port.
PTC9300H features software control system with multiple micro-processors fully integrated to produce highly stable synthesized 3 phase power outputs and reference readings. All control and selection functions are digitally accessible from the PTC 9300H's front control panel or optionally through a RS-232 port to a remote PC. With the proprietary software, testing functions, results and file management are easily performed. The PTC 9300H's VFD illuminated display screen allows for clear readings and menu command operation.
Reference Output Readings
The PTC 9300H's 3 phase synthesized source is capable of simulating broad variations of 3 phase 3 wires and 4 wires networks. These fully rangeable 3 phase power outputs are highly accurate and suitable for laboratory and field applications. With typical accuracy of :
0.02% reading for Voltage, Current and Frequency
0.05% reading for Watts, Vars, Power Factor, WH, VarH
0.015o for Phase Angle
Power Transducer Calibration Function
In transducer calibration function, the provided analog & digital inputs allow the PTC 9300H to perform comparative error calibrations when testing most type of power transducers.
The calibration process is simple as it uses numeric symbol codes in identifying the transducer type, input and output range. A list of commonly used calibration ranges are preprogrammed into PTC 9300H. Some of non-programmed test can be accomplished into test by using programmable input & output ratio function and correcting input and output relationslip to meet requirement.
Calibration is further simplified with the option to perform all tests in either fully automatic or manual steps mode. A high degree of refined calibration is achievable as all test modes are divided into 7 or 11 test points (steps).
Power Meter Calibration Function
Besides transducer calibration, PTC 9300H is also designed for power meter testing & calibration. Similarly, it adopts the simple numeric symbol codes for defining the test meter types, input range and class. Additionally, The power meter calibration function features :
Direct error detection of the test meter
Direct range readings with PT / CT ratio selection
Features
Intelligent Power Instrument
Reference Output Readings
Power Transducer Calibration Function
Power Meter Calibration Function
Model & Ordering
MODEL : PTC 9300H
Specification
3 Phase Reference Output :
Output System :
3 phase 4 wires......Van/Vbn/Vcn......Aa/Ab/Ac
3 phase 3 wires...... Vab/Vbc/Vca...... Aa/Ab/Ac
Voltage Output :
Range selection : Digitally adjustable by numeric keys
Output range :
0~300.00V : L - N phase voltage
0~520.00V : L - L line to line voltage
Resolution : 0.01V
Accuracy :
0.02% (RD) 60~300V range (VL-N)
0.02% (RD) 90~520V range (VL-L)
0.02% (RG) 5~300V range (VL-N)
0.02% (RG) 10~520V range (VL-L)
Temperature influence : 23+ 2oC, 15ppm peroC (5~40oC)
Frequency influence : Typically 0.001% per Hz (48~62Hz range)
Output load : 15VA maximum at 300V, Resistive or inductive impedance or below 2000 PF capacitive impedance
Frequency influence : Typically 0.001% per Hz (48 ~ 62Hz range)
Current Output :
Range selection : Digitally adjustable by numeric keys
Output range : 0~6.2000A, 3 phase
Resolution : 0.0001A
Accuracy :
0.02% RD, 0.25~6.0A range
0.02% RG, 0.01~6.2A range
Temperature influence : 23+2oC, 15ppm peroC (5~40oC)
Frequency influence : Typically 0.001% per Hz (48~62Hz range)
Output load : 15VA maximum at 6 amp Resistive or inductive impedance
Power Output .... Watts and Vars
Frequency influence : Typically 0.002% per Hz (48~62Hz range)
Temperature influence: 23 + 2oC, 30ppm/oC (5~40oC)
Phase Angle :
Range selection : Digitally adjustable by numeric keys
Output range: 0 ... 359.9o
Resolution: 0.1o
Accuracy 0.015o VP3 60V, VL3 90V
A3 0.5A
Temperature influence: 23 + 2oC, 15ppm peroC (5~40oC)
Frequency :
Range : Digitally adjustable 45.00~65.00Hz
Resolution : 0.01Hz
Accuracy: 0.02% RD
Temperature influence: 23 + 2oC, 10ppm peroC (5~40oC)
Transducers Calibration :
Types of measured : V / A / W / Var / Ø / F / WH / VarH
Calibration process : Software control
Selection : Under defined file with code selection
Input ratio : 0~1.2 selectable (initiate to 1)
Output ratio : 0~9.9999 selectable (initiate to 1)
Test mode : Auto or step
Error detection: % RG for analog input, % RD for digital input
Type of calibration : Analog / Digital
Accuracy Test :
Analog input ... for V / A / W / Var / Ø / F types