High Accuracy Reference Reading / 0.05% Reading
The HC 3100 is a new generation of advanced electronics intelligent instruments. Based on digital signal processing (DSP) technology, the HC 3100 is a multi-function precision power meter for three phase systems of 3 or 4 wires and single phase measurements. The measure-ments are comprehensive & fully auto-ranging covering 3 * voltage, 3 * current, active power (KW + KWH), reactive power (KVar + KVarH), apparent power (KVA), frequency, phase angle and power factor. All parameters are of high accuracy up to 0.05% reading.
Energy / Power Calibration Function
HC 3100 also provides energy / power calibration functions with digital input and output, and can be used as a master or sub-reference in measurement. The input is used to accept the measured energy signal when performming calibration function and the output is used as a reference to external comparator or reference standard.
Control Function
VFD illuminated display shows all the electrical measurement parameters in either 5 or 6 digits resolutions. All control and selection functions are simple and accessible with 4 function keys on the front of the instrument.
Features
Reading / display of registered parameters
Measured system (phase and wires)
Energy / power calibration function
Energy / power reference output (pulse)
Conversion time and synchronization
Latch up function for instant reading
Communication via remote control (RS-232)
Model HC 3100 a portable set (wide current input range 0~100A)
Model & Ordering
Model HC 3100 a portable set (wide current input range 0~100A)
Specification
Accuracy of Input Ranges :
Voltage Input :
Range selection : Fully auto-ranging
Vab/Vbc/Vca : 0~600V Line to line voltage, input 1/2 /1.2M
Resolution : 0.001V
Accuracy : 0.05% RD for 50~600V range; 0.05% RG for 6~600V range
Accuracy 0.05% RD, 0.25~10A for 10 range, 5~100A for A100 range; 0.05% RG, 0~0.25A for A100 range, 0~5A for A100 range
Power / Energy :
Range selection : Fully auto-ranging
Power factor : Watts/watthours for cosØ ; vars/varhours for sinØ
Accuracy :
Power Factor :
Range : 0~1.0000 (leading to lagging)
Accuracy : 0.05% RG
Resolution : 0.0001
Phase Angle :
Range : 0.0~359.9o
Accuracy : < 0.03o between 1~359o
Resolution : 0.001o
Frequency :
Range : 45~65 Hz
Accuracy : 0.02% RD
Resolution : 0.0001 Hz
Energy Calibration Function :
Using HC 3100 as a Calibration Reference :
Accuracy : 0.05% RD (refer to energy / power input ranges)
Error detection : % error in reading
Measured input : Digital signals from energy meters Pulse or open collector
Logic of input : Level high3 1.8V, level low2 0.8V
Input range : 0.001 WH to 99.999 WH per pulse or 30~300000 pulses per kWh
Test interval (N) : Numbers of input test pulses, 1~99
Using HC 3100 Pulse Output for External Comparison :
Application : HC 3100 digital output to comparator and other reference standards
Digital output : Refer to HC 3100 digital pulse output
Control and Selection Functions :
System & Conversion :
Phase and wires : 3 phase 4 wires / 3 phase 3 wires / single phase (selected to 3 phase 4 wires)
VStandard : RMS conversion for V, A / real
Conversion : conversion for watts, Mathematic; conversion for other parameters
Applied : Var's measurement is achieved using
Conversion(additional) : sequence shifts (refer to methodology) V to A in 90oC phase, Vab...Ac, Vbc...Aa, Vca...Ab; V to A in 60oC phase Vca...Ac, Vab...Aa (software solution simulating mechanical coordinations)
Reading Display :
Type : illuminated VFD display 4 rows x 32 alphanumeric
Display : Selectable 3 indication display screen; 3 Phase V, A, W, Var, VA, PF, Hz, Ø 3 Phase V, A, W, Var, PF, Hz, EF, ER 3 Phase V, A, W, Var, VA, Hz, W1, W2, W3
Digits : Selectable 5 or 6 digit range resolutions
Terminals / Input : Current Range of 100A terminals or 10 A Terminals
Latch function : Enable or disable
Digital Pulse Output :
Type : Watthour and Varhour pulse outputs
Pulse output selection : Automatic (refer to chart below main manual) or manual
Logic : Level high > 3.6V, level low < 0.5V
Output load : 2K1/2 minimum
Galvanic isolation : 2KV AC RMS
Max. output high frequency 18KHz High Fq. (HF) or Max. output low frequency 1.8 Hz Low Fq. (LF)
Pulse rate chart : Automatic calculation
Formula for calculation of nominal pulse rate, EF
Digital Pulse Input.....Refer to energy calibration function
Conversion Time :
Range (N) : 1~100 cycles settable
Conversion time T : (3N + 2) of input signal periods
Synchronization :
Selection : V synchronized or A synchronized
Latch Function :
Status : Enable / reading latched for instant value